smart HbA1c Test
Summary and Explanation of the Test:
Throughout the circulatory life of the red cell, Haemoglobin A1c (HbA1c) is formed continuously by adduction of glucose to the N-terminal of the haemoglobin beta chain. This process which is non enzymatic reflects the average exposure of haemoglobin to glucose over an extended period. Studies have shown that HbA1c on diabetic subjects was 2-3 fold over levels found in normal individuals.
Method-Range:
Enzymatic HbA1c test: Typical Range (Lot pending!) NGSP: 4,0% - 14,0%
Principal:
Endpoint measurement of the HbA1c- Concentration at 700nm.
HbA1c Test Procedure
1a. Sample preparation
Direct from finger brick

1a.1) Aspirate 20µl whole blood using the smart capillary tube

1a.2) Release filled capillary tube into lysing cup (prefilled with R1b)
1b. Sample preparation
From EDTA tube

1b.1) Aspirate 20µl EDTA whole blood

1b.2) Dispense 20µl sample into lysing cup (prefilled with R1b)
2. Sample preparation

Dispense 2 pump strokes lysing buffer into lysing cup

Shake lysing cup vigorously 6-8 times and wait 6 minutes until complete lysis is evident =Lysate
3. Test application |
|||
![]() |
![]() |
![]() |
![]() |
insert RFID card analyser is reading:
|
pipette 150µl lysate into ERS (Eurolyser-Reagent-System) cuvette | apply patented ERS cap | insert ERS into smart analyser, press "start" - read result |
Test specification
|