Summary
In routine medical diagnostics , the test may be used for the following purposes: To diagnose anemia or polyglobulia, to monitor risk groups for iron deficiency (pregnant women, infants, blood donors, hemodialysis patients, diabetics, athletes, ...), as part of point of care testing when acute bleedings or hemorrhages are suspected, at parasitic diseases as well as quantifying blood loss.
Methode-Measurement Range:
Measuring Range: 0-30 g/dl
Lower Detection Limit: 0,5 g/dl
Principle:
The reagent used binds hemoglobin and its variants to form a coloured complex, which is detected photometrically at 546nm.
HGB Test Procedure
Test kit preparation: Allow single test min. 10 minutes to warm to room temperature by placing the test into the test kit rack. Put test kit package back into refrigerator.
1. Sample collection |
||
| Whole blood from finger tip | Whole blood from primary tube | |
![]() take 20µl whole blood from finger tip |
or
|
![]() take 20µl whole blood (edta) or Serum from a primary tube |
2. Test application |
|||
![]() |
![]() |
![]() |
![]() |
insert RFID card analyser is reading:
|
pipette sample into ERS (Eurolyser-Reagent-System) cuvette | apply patented ERS cap | insert ERS into smart analyser, press "start" - read result |
Test specification
|