CRP test
C-reactive protein
About the parameter
C-reactive protein (CRP) is the best known amongst the acute-phase proteins – a group of proteins whose concentrations increase in the blood as a response to inflammatory disorders. It increases after only 6 hours in acute inflammatory processes associated with bacterial infections, post-operative conditions or tissue damage. Over the course of the disorder its value can rise up to 500 mg/l, which can be quickly and easily determined through POC blood analysis.
The Eurolyser CRP assay is a reliable and easy-to-use quantitative blood test, used to determine c-reactive protein from whole blood as well as from serum. The determination of CRP through blood analysis facilitates the clinical management of patients with infection symptoms.
Point-of-care testing holds the advantage that test results are consulted and interpreted on the spot after only a few minutes. Potentially necessary use of antibiotics can be concluded directly at the doctor’s office – That’s what the Eurolyser instruments were developed for.
Overview
Method
Kinetic determination of the concentration of CRP by photometric measurement at 546 nm or 700 nm of antigen-antibody reaction between antibodies to human CRP bound to polystyrene particles and CRP present in the sample.
Measurement range
This test is an immunoturbidimetric assay with 2 ranges:
Serum:
0.5 – 120 mg/l for 546 nm
1.0 – 120 mg/l for 700 nm
Whole blood:
2.0 – 240 mg/l for 546 nm and 700 nm
(LOT and HCT dependent)
Test duration
Test duration takes approx. 3 minutes measuring serum, approx. 4 minutes measuring whole blood.
Package insert
More details can be found in the package insert and in the package insert for CRP with integrated capillary.
Please note: All package inserts provided on this website are intended for information purposes only! Always process all tests, controls and calibrators strictly according to the package insert provided with the kit.
Appropriate instruments
Evaluation
Assay evaluation
Find the evaluation study for the Eurolyser CRP assay here: Evaluation Eurolyser CRP
Please note: depending on the time of the market launch of an assay, the respective assay evaluation was carried out using either a Eurolyser smart instrument or a Eurolyser CUBE instrument.
Since the measurement method is the same in both instruments (their photometric units are identical in construction), as was proved in this comparison study, all testkit evaluations provided on this website are valid for either instrument!
Additional information
Article numbers
ST 0100 CRP test kit (containing 32 tests)
ST 0102 CRP test kit with integrated capillary (containing 32 tests)
ST 1000 CRP control kit (low/high)
Importance of a haematocrit (HCT) corrected value
Automatic correction of HCT values when measuring CRP facilitates obtaining more precise results. You can read about the reasons in the following PDF document:
Importance of a haematocrit (HCT) corrected value
Eurolyser CRP – Test winner in scientific literature
Two independent researchers from the Netherlands have put eight leading Point-of-care solutions for CRP measurement to the test. Their Article (“Brouwer, N., van Pelt, J. 2015, ‘Validation and evaluation of eight commercially available point of care CRP methods’, Clinica Chimic, no. 439, pp. 195-201) found that the Eurolyser System is the test winner!
- “Taking into account both the analytical validation and the practical evaluation, we consider the Afinion and Smart CRP to be the most suitable analyzers for POCT. Although the correlation coefficient of the Afinion was a bit better than the Smart, Afinion underestimated the CRP concentration by more than 10% and therefore our preferred method is the Smart analyzer.”
(Brouwer & Van Pelt 2015, p.201)
- “A drawback of the Afinion analyzer is that the sample volume is only 1.5 ?L, which means that it is very important to place the cassette directly into the analyzer after adding the blood sample, otherwise an error is generated because the material will clot.”
(Brouwer & Van Pelt 2015, p.198)
- “…the 95% CI for the slope of the linear regression equation compared to the Synchron CRP method included 1.0 only for the Smart analyzer ”
(Brouwer & Van Pelt 2015, p.198)
- “The slope of Smart and AQT90 Flex CRP was around 1.00, and therefore showed the best correlation with the Synchron CRP method.”
(Brouwer & Van Pelt 2015, p.198)
Please note:
The content hoster is very strict in regards to copyright. Therefore we cannot – or rather must not – share the full article on this webpage. If you are interested in reading the whole piece you may follow this link to the content hosters website.
Storage guide
Guidelines for optimal test storage can be found in the following PDF document:
Test Storage Guide (Englisch)
A practical field report
Our distribution partners HITADO conducted an interview with German paediatrician Doctor Ulrich Kuhn about the evaluation of CRP on the Eurolyser CUBE. Read about why he recommends it and how it facilitates his daily operations:
Infection diagnosis on an objective basis – A CRP field report .
Simple procedure – 3 quick steps to get CRP test results with highest accuracy
Features and benefits of Eurolyser CRP test with integrated capillary
Excellent accuracy:
As accurate as a laboratory test, directly traceable to large laboratory systems.
- Quantitative results
- Reproducable CRP results within the 2- 240 mg/l (whole blood) range. This is the broadest range available with any point of care instrument
- Measuring time less than 5 minutes
Simple to use:
Easy to operate for specialists as well as for non-laboratory staff
- Simply performed on a finger prick blood sample (alternatively: venous blood or serum)
- No pipetting steps necessary
- Built-in calibration
- Instrument with state of the art LED photometer technology
- 100 patient data storage
- LIS or LIM connectivity
Summary
Euroylser CRP is a simple test for quantitative determination of C-reactive protein (CRP) in whole blood or serum.
C-reactive protein (CRP) is the best known among the acute-phase proteins, a group of proteins whose concentration increases in blood as a response to inflammatory disorders. It increases already after 6 hours in acute inflammatory processes associated with bacterial infections, post operative conditions or tissue damage.
When the test is performed at the point of care the result will be available during patient consultation and can effectively guide antibiotic use. The Eurolyser system – consisting of a small instrument and a ready-to-use kit – is specifically designed for use in primary health care settings and is instantly ready for operation, providing immediate test results.